Literature review · started Nov 18, 2026
Lung cancer screening with low-dose CT
Example This is an example review. You can read every page, but not change it. The AI screen in this example ran on a sample of 23 of the 9,965 studies found, so most studies show as not screened.
Low dose megavoltage cone beam computed tomography with an unflattened 4 MV beam from a carbon target
Not decided
Not in the sample the AI screened
- Decided by
- Nobody yet
- Found in
- Crossref
- Free full text
- Not looked for in this run
Abstract
Megavoltage cone beam computed tomography (MVCBCT) is routinely used for visualizing anatomical structures and implanted fiducials for patient positioning in radiotherapy. MVCBCT using a 6 MV treatment beam with high atomic number target and flattening filter in the beamline, as done conventionally, has lower image quality than can be achieved with a MV beam due to heavy filtration of the low‐energy bremsstrahlung. The unflattened beam of a low target has an abundance of diagnostic energy photons, detected with modern flat panel detectors with much higher efficiency given the same dose to the patient. This principle guided the development of a new megavoltage imaging beamline (IBL) for a commercial radiotherapy linear accelerator. A carbon target was placed in one of the electron primary scattering foil slots on the target‐foil slide. A PROM on a function controller board was programed to put the carbon target in place for MVCBCT. A low accelerating potential of 4.2 MV was used for the IBL to restrict leakage of primary electrons through the target such that dose from x rays dominated the signal in the monitor chamber and the patient surface dose. Results from phantom and cadaver images demonstrated that the IBL had much improved image quality over the treatment beam. For similar imaging dose, the IBL improved the contrast‐to‐noise ratio by as much as a factor of 3 in soft tissue over that of the treatment beam. The IBL increased the spatial resolution by about a factor of 2, allowing the visualization of finer anatomical details. Images of the cadaver contained useful information with doses as low as 1 cGy. The IBL may be installed on certain models of linear accelerators without mechanical modification and results in significant improvement in the image quality with the same dose, or images of the same quality with less than one‐third of the dose.
From Crossref.
The criteria
-
Include if: The document is in English.
Meets (checked by an automatic rule)What the rule found: language_in: language=en; title_words=16; title_english_function_words=4; title_english_share=0.2500; title_reads=english; abstract_words=305; abstract_english_function_words=125; abstract_english_share=0.4098; abstract_reads=english; language_check=metadata_listed