Lung: Feasibility of a Method for Changing Tube Current during Low-Dose Helical CT

Shigeki Itoh; Mitsuru Ikeda; Yoshine Mori; Koujiro Suzuki; Akiko Sawaki; Shingo Iwano; Hiroko Satake; Shoji Arahata; T Isomura; Masahiro Ozaki; Takeo Ishigaki (2002). Radiology

DOI 10.1148/radiol.2243010874

Not decided

Not in the sample the AI screened

Decided by
Nobody yet
Found in
OpenAlex
Free full text
Not looked for in this run

Abstract

A method for changing the tube current during helical scanning was applied to low-dose computed tomography (CT) in the lung. The changing method resulted in significant equalization of image noise in various lung sections compared with that at scanning with constant tube current. Detectability of nodules was equivalent between 60 mA and the changing method, whereas degradation occurred at 20 mA. This method seems feasible for the low-dose CT of lung cancer screening.